Designs and validates quantitative targeted metabolomics assays (MRM/SRM on triple-quadrupole, PRM on high-resolution instruments) to report absolute concentrations. Covers the internal-standard strategy (external cal -> global IS -> standard addition -> stable-isotope-labeled IS), weighted calibration judged by back-calculated %RE not R-squared, ion-ratio quantifier/qualifier confirmation, matrix-effect/recovery characterization, and ICH M10 method validation. Use when quantifying a closed panel of known metabolites with units, building or validating an LC-MS/MS assay, choosing an IS or calibration weighting, or judging whether a reported concentration is trustworthy. For untargeted feature detection see metabolomics/xcms-preprocessing; for group statistics see metabolomics/statistical-analysis; for flux/MID/tracing see metabolomics/isotope-tracing.

General

What this skill does


## Version Compatibility

Reference examples tested with: R 4.3+, ggplot2 3.5+, Skyline 23.1+, pandas 2.2+, numpy 1.26+

Before using code patterns, verify installed versions match. If versions differ:
- R: `packageVersion('<pkg>')` then `?function_name` to verify parameters
- Python: `pip show <package>` then `help(module.function)` to check signatures
- CLI: `<tool> --version` then `<tool> --help` to confirm flags

An absolute concentration requires three inputs the code cannot supply: an authentic reference standard (its certificate-of-analysis purity scales every reported number), a stable-isotope-labeled internal standard that co-elutes with the analyte, and a per-analyte validation record. Without these, the workflow below produces relative peak-area ratios dressed as concentrations.

If code throws ImportError, AttributeError, or TypeError, introspect the installed
package and adapt the example to match the actual API rather than retrying.

# Targeted Metabolomics Analysis

**"Quantify these specific metabolites and give me concentrations with units"** -> Fit weighted calibration curves on authentic standards, normalize each analyte to a co-eluting stable-isotope-labeled internal standard, confirm identity by ion ratio, and report concentrations only within the validated range.
- CLI: Skyline builds the small-molecule transition list, integrates peaks, fits weighted calibration, and exports via the Document Grid.
- R / Python: post-export curve fitting, back-calculated %RE checks, IS normalization, ion-ratio confirmation, validation metrics.

## The Single Most Important Modern Insight -- A Concentration Is a Chain of Cancellations, and Matrix Effects Are the Term That Fails to Cancel

Ion suppression is competition for charge and droplet surface in the electrospray source: a co-eluting matrix component (phospholipids late in a reversed-phase gradient, salts at the void) steals ionization from the analyte. Suppression is a property of the co-elution, not of the analyte, so it is retention-time-dependent and lot-dependent. The only mechanism that truly removes it is a stable-isotope-labeled internal standard (SIL-IS) that sits in the identical droplet at the identical instant: the suppression cancels in the analyte/IS area ratio. An IS that elutes even half a minute away samples a different point on the suppression landscape and injects new error rather than removing it. Every other safeguard in this skill -- weighting, ion ratios, validation -- assumes this cancellation is working; the gap between a solvent calibration curve and a matrix-matched curve is a direct readout of how badly the IS is failing.

## Targeted vs Untargeted -- Different Experiments, Not Two Settings

| Axis | Untargeted (discovery) | Targeted (quantification) |
|---|---|---|
| Analyte set | Open -- everything ionizable | Closed -- a panel defined before acquisition |
| Output | Relative fold-change; often putative IDs | Absolute concentration for confirmed analytes |
| Instrument | High-res full-scan / DDA (Orbitrap, QTOF) | Triple-quad SRM/MRM, or high-res PRM |
| Validation | QA/QC framework (Broadhurst, mQACC) | Full bioanalytical validation possible (ICH M10) |
| Question | "What changed?" | "How much is there?" |

Targeted buys sensitivity and absolute quant by spending scope (only what is on the list is seen) and up-front method development. Common pattern: untargeted discovery -> targeted validation of the hits. Feature detection upstream is metabolomics/xcms-preprocessing; this skill begins once the panel and transitions are defined.

## Acquisition Mechanics -- MRM/SRM and PRM

A transition is a precursor-m/z -> product-m/z pair plus a tuned collision energy. On a triple quadrupole, Q1 isolates the precursor, the collision cell fragments it, Q3 isolates one product: the double mass filter is the source of MRM sensitivity. SRM monitors one transition; MRM multiplexes many. Each analyte should carry at least two transitions -- a quantifier (most intense/cleanest, used for concentration) and one or more qualifiers (orthogonal confirmation). PRM replaces Q3 with a high-resolution analyzer that records the full product spectrum in parallel, so transitions are chosen post hoc and isobaric interferences are resolved by exact mass; MRM still wins on absolute sensitivity and very large panels. Dwell time is the signal-accumulation time per transition; cycle time must stay short enough for at least 10-15 points across each chromatographic peak (convention). Scheduled MRM monitors each transition only within a retention-time window so a large panel keeps adequate dwell -- but a peak that drifts out of its window vanishes with no error message, the classic scheduled-MRM failure.

## Decision Tree -- Quant Goal -> IS + Calibration + Validation Depth

| Goal / situation | Internal standard | Calibration | Validation depth | Why |
|---|---|---|---|---|
| Clinical / regulated / PK number | One SIL-IS per analyte (13C/15N) | Multi-level weighted curve, judged by %RE | Full ICH M10 (accuracy, precision, MF, recovery, carryover, stability, ISR) | A number driving a decision must carry its evidence |
| Cross-study quantitative claim | SIL-IS per analyte or per RT/chemical cluster | Multi-level weighted | Accuracy/precision + matrix-factor on QCs | Comparability across runs demands characterized bias |
| Exploratory research, ranking | Few global IS, or per-class | 1/x^2 weighted, low-end %RE checked | Broadhurst/mQACC QC discipline (pooled QC, blanks, RSD filtering) | Relative comparison tolerates residual matrix bias |
| Dirty matrix, isobaric interferences | SIL-IS + high-res | PRM, post-hoc transitions | Selectivity dominated | Exact-mass product resolves co-eluters a unit-resolution Q3 cannot |
| Large standardized panel (600+) | Kit-supplied class IS | Single/limited-point (vendor) | Vendor + bridging study before pooling sites | Kit buys comparability and throughput, not per-analyte full-validation accuracy |
| Carbon source / pathway rate | (tracer, not IS) | -- | -- | Flux question: hand off to metabolomics/isotope-tracing; MID measures rate, not pool size |

The IS rule of thumb: ask how far (in retention time and chemistry) each analyte is from its assigned IS -- that distance is the size of the uncorrected matrix error. 13C/15N at non-exchangeable positions are preferred over deuterium: deuterium causes a small reversed-phase retention shift (the deuterium isotope effect) that can chromatographically separate the IS from its analyte so it stops correcting suppression, and labile deuteriums back-exchange to H. If forced to a deuterated IS, verify co-elution by overlaying analyte and IS chromatograms.

## Calibration and Weighting

| Weighting | When | Effect |
|---|---|---|
| Unweighted (OLS) | Narrow range, near-constant variance | High points dominate; low-end bias on heteroscedastic MS data -- usually wrong |
| 1/x | Moderate range (1-2 orders) | Down-weights high concentrations; restores low-end fit |
| 1/x^2 | Wide range (3+ orders), the common LC-MS default | Aggressively down-weights the top; can over-weight the low end -- still compare, do not reflex |
| Quadratic | Genuine, mechanism-explained curvature (detector saturation) | Never to paper over a bad linear fit |

MS detector response is heteroscedastic -- absolute variance grows with concentration -- so weighting models the variance structure (1/x and 1/x^2 are parametric stand-ins for 1/variance). Select empirically: fit candidate weightings, then pick the one minimizing the sum of absolute back-calculated relative error (%RE) across levels, especially the bottom two or three. R-squared is the wrong instrument: it is dominated by high-leverage top points, so a curve with R-squared 0.999 can be +40% biased at the LLOQ. Use a fitted (non-zero) intercept; forcing the line through the origin re-introduces low-end bias. The blank (matrix only) and zero (matrix + IS) are diagnostic, not calibration points.